KSL-1800X-KA-UL is a bench-top ultra-high temperature muffle furnace built by Kanthal HT (1900 grade) MoSi2 heating elements with Max working temperature of 1800°C. It is ready to pass all safety tests worldwide.
SPECIFICATIONS:
Furnace Structure
|
- Double shell steel case with a cooling fan to keep outside surface cool
- High purity alumina fiber insulation for max. energy saving
- ZrO2 ceramic liner can stand at working temperature up to 2200°C
- All electric components are UL or Met recognized and ready to pass any safety test worldwide
|
| Power |
- AC 208V-240V single phase, 50/60 Hz (50A air breaker required)
- 4.8 KW
- 10 feet long 10-3 AWG heavy duty UL approved power cable is included
- Extra length cable can be ordered by click the picture below
- The plug is not included. You can order from your local hardware store according to the female plug in your lab, or order from the link below (30A plug)
|
| Working Temperature |
- 1800° C (< 1 hour)
- 1750° C (continuous)
|
| Heating Rate |
- 0~10°C / min (suggestion =< 5°C)
|
|
Temperature Uniformity
|
|
|
Heating Elements
|
Kanthal Super-1900 Grade MoSi2, 4 pcs
|
|
Thermocouple
|
B-Type (Pt-Rh to Pt-Rh) with Alumina tube
|
Temperature controller
|
- Eurotherm 3000 Series Programmable Temperature Controller.
- 1 program, 24 segments programmed with ramping, cooling, and dwelling steps
- Temperature Accuracy: ± 0.1ºC
- RS485 port is built-in for PC operation
- Operation software is included for PC operation.
- Brand new laptop with Microsoft Windows 10 with pre-installed software is available upon request.
|
| Compliance |
- CE Certified
- NRTL or CSA certification is available at the extra cost
|
| Application Notes |
- For longer service life, we recommend the heating rate not to exceed 10 °C /min.
- The cooling rate shall also not exceed 10 °C /min.
- This furnace lacks vacuum sealing, toxic or explosive gases are not recommended for use with this furnace without necessary safety controls and supervision.
- Small cracks may appear on the surface of the refractory ceramics over extended use. This is a normal occurrence and the cracks may be repaired with alumina coating.
- Please click here to learn how to DIY a gas purging system.
- The refractory block must be inserted before closing the door.
|
STANDARD PACKAGE:
| P/N |
Descriptions |
Quantity
|
Picture |
| KSL-1700X |
KSL-1800X-UL High-Temperature Muffle Furnace |
1 set |
 |
| KSL7-01 |
Kanthal Super-1900 Grade MoSi2, 4 pcs, |
4 pcs |
|
| KSL7-02 |
B-Type Pt-Rh to Pt-Rh Thermocouples with Alumina Protection Tube |
1 set |
|
| KSL7-03 |
Al2O3 Sample Plate ($95.00/ea) |
1 pcs |
|
| KSL7-04 |
Al2O3 Furnace Door Block ($180.00/ea) |
1 pcs |
|
| KSL7-06 |
Crucible Clip ($25.00/ea) |
1 pair |
|
| KSL7-07 |
Operation Manual |
1 set |
|